Chem Sity GPS
Practice

Real Chemistry Lab

Measurement, Accuracy, and Precision

Real data from a volumetric-glassware lab. Each block is labeled so you can tell what was recorded at the bench, what was calculated from it, and what is explanation.

Recorded observationCalculationEducational explanation

Part A · Pipette verification

Recorded observation

Recorded masses and readings

10 mL beaker

Empty mass
31.0163 g
Filled mass
38.6512 g
Water mass
7.6349 g

Volumetric pipette

Delivered water mass
9.9723 g

Graduated cylinder

Empty mass
39.7026 g
Filled mass
49.5116 g
Water mass
9.8090 g

Buret

Delivered water mass
10.0324 g
Initial reading
0.01 mL
Water temperature
25 °C

Water density used for the documented buret calculation: 0.99705 g/mL (at the recorded 25 °C).

Calculation

Buret: delivered volume and error

  1. Delivered volume10.0324 g ÷ 0.99705 g/mL = 10.0621 mL
  2. Signed volume error10.0621 mL − 10.0000 mL = +0.0621 mL
  3. Absolute error|+0.0621 mL| = 0.0621 mL
  4. Signed percent error(+0.0621 mL ÷ 10.0000 mL) × 100 = +0.621%

The positive sign means the buret delivered slightly more than the nominal 10.0000 mL.

Not established by this data

  • A complete accuracy ranking of all five devices was not conclusively established.
  • The final buret reading was not recorded.
  • Temperatures for the other devices were not recorded, so their masses are not converted to volumes here.
  • Micropipette values from Part A were unclear and are not reported.

Part B · Precision comparison

Recorded observation

Recorded incremental masses (g)

Five recorded incremental masses for each device, in grams
Device12345
Volumetric pipette0.98430.98690.97480.98430.9898
P1000 micropipette1.00620.97860.97410.97920.9896
Calculation

Summary statistics

Volumetric pipette

Mean mass
0.98402 g
Sample standard deviation
≈ 0.00563 g
Mass %RSD
≈ 0.572%

P1000 micropipette

Mean mass
0.98554 g
Sample standard deviation
≈ 0.01287 g
Mass %RSD
≈ 1.306%
Educational explanation

What these numbers do and do not show

The volumetric pipette had the lower %RSD, so it showed better observed mass-delivery repeatability (higher precision) in this dataset.

This does not prove greater volume accuracy. The Part B water temperatures were not recorded, so the masses cannot be reliably converted to volumes and compared with nominal values.

Key concepts

Educational explanation

Measurement vocabulary

Mass
The amount of matter in a sample, measured on a balance in grams.
Density
Mass per unit volume (g/mL). Water density changes with temperature: 0.99705 g/mL at 25 °C.
Delivered volume
Mass of water delivered ÷ water density at the recorded temperature.
Signed error
Measured − expected. The sign shows whether a device delivered more (+) or less (−) than nominal.
Absolute error
The size of the error without its sign.
Percent error
Error ÷ expected value × 100. A measure of accuracy.
Sample standard deviation
The spread of repeated measurements around their mean, using n − 1 in the denominator.
%RSD
Standard deviation ÷ mean × 100. A measure of precision relative to the average.

Scientific integrity and corrective action

Recorded observation

Day 1: equipment malfunction

The original balance showed unstable readings, sometimes changing by 100 g or more. The experiment was stopped. A replacement Mettler balance was located and tested in Room 466.

Recorded observation

Day 2: measurements completed

The replacement balance was used to complete the measurements.

Educational explanation

Why this matters

  • Stop when equipment malfunctions. More readings from a faulty instrument only produce more invalid data.
  • Document the problem in the lab record, even if results look reasonable.
  • Take corrective action: replace the instrument and test it before use.
  • Separate reliable from invalid data. Report only measurements taken with verified equipment.

Check your understanding

0 of 6 solved

Lab question 1

Which statement correctly distinguishes accuracy from precision?

Choose an answer

Lab question 2

The buret delivered 10.0324 g of water. Using the documented water density of 0.99705 g/mL, what volume was delivered?

Choose an answer

Lab question 3

The buret was used to deliver a nominal 10.0000 mL, and the calculated delivered volume was 10.0621 mL. What is the signed percent error?

Choose an answer

Lab question 4

In Part B, the volumetric pipette had a mass %RSD of about 0.572%, and the P1000 micropipette about 1.306%. Which conclusion does this dataset support?

Choose an answer

Lab question 5

Why must water temperature be recorded when converting a measured mass of water to volume?

Choose an answer

Lab question 6

On Day 1, the balance gave unstable readings, sometimes changing by 100 g or more. What was the correct scientific response?

Choose an answer